Skip to content
  • Home
  • Knowledge
  • About
  • Contact
  • Privacy Policy
  • Home
  • Knowledge
  • About
  • Contact
  • Privacy Policy

Knowledge Home

1
  • Guidelines for Wiki Corrections

Wiki

1
  • Wiki Home

Case Studies

3
  • Case Studies Home
  • PDA
  • Dentistry with comorbidities

Repo Lab

3
  • AI Workflow and Considerations
  • Model Evaluation
  • Repo Lab Home

Cardiovascular

9
  • Intravenous Fluid Rate Selection During Anesthesia for Dogs and Cats
  • Goal-Directed Fluid Therapy in Veterinary Patients
  • Interpretation of Arterial Pressure Tracings During Anesthesia
  • Pressure Waveform Analysis and Plethysmography for Preload Assessment in Anesthetized Animals
  • Subaortic Stenosis in Dogs
  • Feline Hypertrophic Cardiomyopathy
  • Mitral Valve Disease in Dogs and Cats
  • Coagulation and Hemostasis
  • Cardiovascular Physiologic Parmaters

Commerical LLMs

4
  • Why Most AI Chatbots Are Poor Sources of Medical Advice for Professionals
  • OpenAI
  • Claude
  • Commercial LLMs

Data Science

3
  • Causal Quartets
  • Favorite DS Podcasts
  • Data Science

Equipment

5
  • Thermal Support Devices for Anesthetized Dogs and Cats
  • Inhalant Anesthetic Vaporizers in Veterinary Medicine
  • Endotracheal Tube
  • Laryngoscope
  • Equipment

Bayesian Statistics

8
  • Weight Loss, Adaptation and Other Asymmetric Biological Phenomena
  • Statistical Paradoxes and Ignorant People
  • Learning Bayesian Statistics
  • Statistical Rethinking
  • BDA3
  • Aubry Clayton’s Bernoulli’s Fallacy
  • E.T. Jaynes’ Probability Theory: The Logic of Science
  • Bayesian Statistics

Monitoring

6
  • Artifacts in End-Tidal CO2 Monitoring and Capnography in Dogs and Cats
  • Body Temperature
  • Depth of Anesthesia
  • Respiration and Ventilation
  • Arterial Blood Pressure
  • Overview

Automated Workflow

2
  • n8n
  • Automated Workflow

Procedure Specifics

2
  • Bronchoscopy in Dogs and Cats
  • Considerations for Veterinary Anesthetists When Working Around MRI

Pathophysiology

5
  • Pathophysiology of Sepsis and Shock in Dogs and Cats
  • Pathophysiology of Aspiration Pneumonia
  • Chronic Kidney Disease
  • PDA
  • Overview

RAG

2
  • Vector Store Database
  • RAG

Pharmacology

19
  • Commonly Used CRI Drugs in Veterinary Anesthesia: A Reference Guide
  • Reversal of Neuromuscular Junction Blockers in Dogs and Cats
  • Considerations for Selecting Induction Drugs
  • Opioids in Veterinary Anesthesia: A Summary
  • Pharmacology of Fentanyl in Dogs and Cats
  • Buprenorphine
  • Clinical Pharmacology of Methadone in Dogs and Cats
  • Opinion-Why Midazolam Sucks as a Co-induction Agent with Propofol
  • Historical Perspective: Benzodiazepines in Co-Induction with Ketamine and Propofol
  • Atropine vs. Glycopyrrolate
  • Drug-Drug Interactions and Polypharmacy
  • Norepinephrine During Anesthesia in Dogs and Cats
  • Dopamine vs Dobutamine: Pharmacological Comparison
  • Dexmedetomidine
  • Buprenorphine
  • Alfaxalone
  • Isoflurane
  • Propofol
  • Atropine

GitHub

1
  • GitHub

Endocrine

3
  • Addison’s Disease
  • Diabetes Mellitus in Dogs and Cats
  • Endocrine

Hugging Face

1
  • Hugging Face

Nervous System

4
  • Seizures (Idiopathic Epilepsy)
  • Post-anesthetic Sensory Deficit
  • Anesthetic Actions on CNS
  • Central Nervous System Overview

Local Hosted LLMs

3
  • PydanticAI
  • Ollama
  • Local Hosted LLMs

Hepatorenal

3
  • Anesthetic Considerations for Patients with Protein-Losing Nephropathy
  • Anesthetic Management for Cats and Dogs with Hepatic Shunts
  • Liver and Kidney Overview

Respiratory

6
  • Mechanical Ventilation in Dogs and Cats: A Comprehensive Summary
  • Preoxygenation Before Anesthesia in Dogs and Cats: Advantages and Disadvantages
  • Feline Asthma
  • Laryngeal Paralysis
  • Brachycephalic Airway
  • Comparative Physiologic Parameters

Other

10
  • Navigating the Legal Complexities of Extralabel Drug Use in Veterinary Medicine
  • When to Use Continuous Rate Infusions for Medication Delivery: A Pharmacoeconomic Analysis
  • Using AI Chatbots to Calculate Veterinary Medication Dosages: Fentanyl CRIs Made Simple
  • Managing Esophageal Reflux During Canine Anesthesia
  • Supervision of Non-Veterinarians Delivering Anesthesia
  • Learning Veterinary Anesthesia Skills
  • The Glycocalyx: Structure and Significance
  • The Limitations of Mortality Rate as an Anesthesia Safety Indicator
  • The Value of Monitoring Guidelines in Anesthesia Practice
  • The Pros and Cons of Using Anesthesia Checklists in Veterinary Medicine
View Categories
  • Home
  • Docs
  • Knowledge Home
  • Wiki
  • Monitoring
  • Respiration and Ventilation

Respiration and Ventilation

3 min read

Methods for Monitoring Respiration and Ventilation in Veterinary Patients #

Respiratory monitoring in veterinary medicine is essential for assessing oxygenation, ventilation, and respiratory mechanics. Various techniques exist, ranging from basic observation to advanced electronic monitoring, each with specific applications, limitations, and sources of error.


1. Physical Assessment (Observation & Auscultation) #

1.1 Methodology #

  • Direct observation of chest movement, nostril flare, or abdominal effort.
  • Palpation of the thorax for paradoxical breathing patterns.
  • Auscultation using a stethoscope to detect breath sounds and abnormalities.

1.2 Equipment Used #

  • Stethoscope for lung sound assessment.
  • Respiratory rate monitor (manual counting of breaths per minute).

1.3 Errors and Artifacts #

  • Observer bias or human error in counting.
  • Panting or shivering can obscure accurate respiratory rate assessment.
  • Auscultation limitations (e.g., muffled lung sounds due to obesity or pleural effusion).

1.4 Normal Ranges #

  • Dogs: 10–30 breaths per minute (bpm).
  • Cats: 20–40 bpm.

1.5 Supporting References #

  • Silverstein D.C., Hopper K. (2015). Small Animal Critical Care Medicine.

2. Pulse Oximetry (SpO₂ Monitoring) #

2.1 Methodology #

  • Measures oxygen saturation (SpO₂) of hemoglobin in arterial blood.
  • Uses infrared light absorption to estimate SpO₂ levels.

2.2 Equipment Used #

  • Pulse oximeter sensor (placed on the tongue, ear, or toe pad).
  • Handheld or multiparameter monitor for real-time readings.

2.3 Errors and Artifacts #

  • Motion artifacts (patient movement can disrupt signal).
  • Hypoperfusion (shock, vasoconstriction) may cause unreliable readings.
  • Pigmentation or thick tissues (dark skin or excessive fur may interfere).

2.4 Normal Ranges #

  • Acceptable SpO₂: ≥92% (room air), <90% suggests hypoxemia.

2.5 Supporting References #

  • Hofmeister E.H., et al. (2010). “Accuracy of Pulse Oximeters in Dogs and Cats.” J Vet Emerg Crit Care.

3. Capnography (End-Tidal CO₂ Monitoring) #

3.1 Methodology #

  • Measures end-tidal CO₂ (EtCO₂) to assess ventilation effectiveness.
  • Provides waveform analysis (capnogram) reflecting respiratory patterns.

3.2 Equipment Used #

  • Mainstream or sidestream capnograph (attached to an endotracheal tube).

3.3 Errors and Artifacts #

  • Leakage around the ET tube causes falsely low EtCO₂ values.
  • Moisture buildup in the sample line can interfere with measurements.
  • Obstructed airway (e.g., mucus plug) may create abnormal waveforms.
  • Non-rebreathing systems may dilute expired gases resulting in deformed waveforms and lower readings

3.4 Normal Ranges #

  • Dogs & Cats: EtCO₂: 35–45 mmHg.

3.5 Supporting References #

  • Steffey, E.P., et al. (2003). “Capnography in Veterinary Anesthesia.” Vet Anesth Analg.

4. Blood Gas Analysis (PaO₂ & PaCO₂ Monitoring) #

4.1 Methodology #

  • Arterial blood sample analysis for oxygen (PaO₂) and carbon dioxide (PaCO₂) levels.
  • Determines acid-base balance via pH, bicarbonate (HCO₃⁻), and lactate.

4.2 Equipment Used #

  • Blood gas analyzer (e.g., i-STAT, VetStat).
  • Heparinized arterial blood sample (from femoral, dorsal pedal, or auricular artery).

4.3 Errors and Artifacts #

  • Air bubbles in the sample can falsely increase PaO₂.
  • Delayed analysis (>10 min) leads to metabolic changes affecting accuracy.

4.4 Normal Ranges #

  • PaO₂ (room air): 90–110 mmHg.
  • PaCO₂: 35–45 mmHg.
  • pH: 7.35–7.45.

4.5 Supporting References #

  • Hopper K., et al. (2014). “Interpretation of Blood Gas Results in Small Animals.” J Vet Emerg Crit Care.

5. Apnea and Respiratory Effort Monitoring #

5.1 Methodology #

  • Monitors respiratory effort and pauses (apnea episodes) in anesthetized patients.

5.2 Equipment Used #

  • Apnea alarm sensors (detect chest wall movement).
  • Esophageal pressure monitors (measure inspiratory effort).

5.3 Errors and Artifacts #

  • False alarms due to patient movement.
  • Sensor misplacement can lead to missed apnea detection.

5.4 Normal Ranges #

  • Apnea detection threshold: ≥10 seconds without spontaneous breath.

5.5 Supporting References #

  • Kelmer, E., et al. (2016). “Apnea Detection in Anesthetized Dogs and Cats.” Vet Anesth Analg.

6. Respiratory Mechanics (Tidal Volume & Compliance Monitoring) #

6.1 Methodology #

  • Measures tidal volume (Vt), airway resistance, and lung compliance during mechanical ventilation.

6.2 Equipment Used #

  • Anesthesia ventilator with built-in monitoring.
  • Spirometer for measuring lung volumes.

6.3 Errors and Artifacts #

  • Leakage around the ET tube affects volume measurements.
  • Incorrect ventilator settings may lead to barotrauma or atelectasis.

6.4 Normal Ranges #

  • Tidal volume: 10–15 mL/kg.

6.5 Supporting References #

  • Bateman, L., et al. (2017). “Ventilatory Management in Veterinary Patients.” J Vet Emerg Crit Care.

Comparison of Methods #

MethodMeasuresBest Use CaseKey Limitation
Observation & AuscultationRate, depth, effortRoutine exam, emergency triageSubjective, inaccurate in critical cases
Pulse OximetrySpO₂, pulse rateAnesthesia, ICU monitoringAffected by perfusion, pigmentation
CapnographyEtCO₂, ventilationAnesthesia, mechanical ventilationTube leaks, moisture interference
Blood Gas AnalysisPaO₂, PaCO₂, pHCritical care, respiratory distressRequires arterial sampling
Apnea MonitoringRespiratory effortAnesthesia, post-op recoveryFalse alarms, sensor misplacement
Respiratory MechanicsVt, complianceMechanical ventilationRequires specialized equipment

Conclusion #

Respiratory monitoring is essential in veterinary medicine for ensuring adequate oxygenation and ventilation. Selection of a monitoring technique should be based on clinical context, patient condition, and available equipment. Advanced methods like capnography and blood gas analysis provide the most precise data, while basic observation and pulse oximetry remain valuable for general monitoring.


References #

  1. Silverstein D.C., Hopper K. (2015). Small Animal Critical Care Medicine.
  2. Hofmeister E.H., et al. (2010). “Accuracy of Pulse Oximeters in Dogs and Cats.” J Vet Emerg Crit Care.
  3. Steffey, E.P., et al. (2003). “Capnography in Veterinary Anesthesia.” Vet Anesth Analg.
  4. Hopper K., et al. (2014). “Interpretation of Blood Gas Results in Small Animals.” J Vet Emerg Crit Care.
  5. Kelmer, E., et al. (2016). “Apnea Detection in Anesthetized Dogs and Cats.” Vet Anesth Analg.
  6. Bateman, L., et al. (2017). “Ventilatory Management in Veterinary Patients.” J Vet Emerg Crit Care.

This outline provides a comprehensive, evidence-based reference for veterinary respiratory monitoring. 🐾💨

Updated on February 24, 2025

What are your Feelings

  • Happy
  • Normal
  • Sad
Depth of AnesthesiaArterial Blood Pressure

Powered by BetterDocs

Table of Contents
  • Methods for Monitoring Respiration and Ventilation in Veterinary Patients
  • 1. Physical Assessment (Observation & Auscultation)
  • 1.1 Methodology
  • 1.2 Equipment Used
  • 1.3 Errors and Artifacts
  • 1.4 Normal Ranges
  • 1.5 Supporting References
  • 2. Pulse Oximetry (SpO₂ Monitoring)
  • 2.1 Methodology
  • 2.2 Equipment Used
  • 2.3 Errors and Artifacts
  • 2.4 Normal Ranges
  • 2.5 Supporting References
  • 3. Capnography (End-Tidal CO₂ Monitoring)
  • 3.1 Methodology
  • 3.2 Equipment Used
  • 3.3 Errors and Artifacts
  • 3.4 Normal Ranges
  • 3.5 Supporting References
  • 4. Blood Gas Analysis (PaO₂ & PaCO₂ Monitoring)
  • 4.1 Methodology
  • 4.2 Equipment Used
  • 4.3 Errors and Artifacts
  • 4.4 Normal Ranges
  • 4.5 Supporting References
  • 5. Apnea and Respiratory Effort Monitoring
  • 5.1 Methodology
  • 5.2 Equipment Used
  • 5.3 Errors and Artifacts
  • 5.4 Normal Ranges
  • 5.5 Supporting References
  • 6. Respiratory Mechanics (Tidal Volume & Compliance Monitoring)
  • 6.1 Methodology
  • 6.2 Equipment Used
  • 6.3 Errors and Artifacts
  • 6.4 Normal Ranges
  • 6.5 Supporting References
  • Comparison of Methods
  • Conclusion
  • References